Multilogger – thermo-hygro-CO2-barometer with 2 MiniDIN and 2 Terminals

SKU: C95M1323 Categories: , , Tags: , , , , , , , , Brand:
BRAND
M1323

Battery powered datalogger with built-in CO2 and atmospheric pressure sensor and 4 universal inputs for measurement of temperature, humidity, 0-10V voltage, 0-20mA current, two-state signal and for counting pulses.

Multilogger can be installed permanently to Ethernet network or work as portable device.

Included is traceable calibration certificate in accordance with EN ISO/IEC17025.

The Multilogger is a battery operated datalogger capable of measuring and recording sensor and electrical values through its inputs.

Datalogger can be connected to Ethernet. Then measured values can be viewed on a web browser, as well as stored and downloaded for later analysis.

 carbon dioxide

Extended measuring range available for an additional charge

  • Measuring range: 0 to 10 000 ppm
  • Accuracy: 100 ppm + 5% of the measured value at 25 °C and 1013 hPa

Technical data

Measured value


Temperature

Relative humidity

Counted values (dew point…)

CO2

Atmospheric pressure

Current

Voltage

Two-state signal

Pulses

TEMPERATURE SENSOR – external probe Pt1000

Measuring range
-200 to +600 °C
Accuracy
±0.2 °C (-200 to +100 °C);
±0.2 % from reading from +100 to +600 °C
Resolution
0.1 °C

TEMPERATURE+HUMIDITY SENSOR – external probe T+RH DIGI

Measuring range
according to the connected probe Digi/M
Accuracy
according to the connected probe Digi/M
Resolution
0.1 °C

DEW POINT

Measuring range
according to the connected probe Digi/M
Accuracy
according to the connected probe Digi/M
Resolution
0.1 °C

ATMOSPHERIC PRESSURE SENSOR

Measuring range
700 hPa to 1100 hPa
Accuracy
±1,3 hPa at 23 °C
Resolution
0.1 hPa

CO2 SENSOR

Measuring range
0 to 2 000 ppm
Accuracy
50 ppm + 2 % of the measured value at 23 °C a 1013 hPa
Resolution
1 ppm
DC VOLTAGE
Measuring range
0 to 10 V
Accuracy
±10 mV
Resolution
better than 100 µV

DC CURRENT

Measuring range
0 to 20 mA
Accuracy
±20 µA
Resolution
better than 0,1 µA

BINARY SIGNAL

Input voltage for level “switched on”/
<0,4 V/> 2 V
Minimum/maximum input voltage
0 V/+30 V DC
I mpedance of the “switched on”/
< 10 kOhm/> 300 kOhm

COUNTER INPUT

Range of counter
24 bit (16 777 215) the counter overflow can be allowed
Maximum input frequency
200 Hz

GENERAL TECHNICAL DATA

Operating temperature
-10 to +60 °C
Channels
internal atmospheric pressure sensor a CO2, 2x external temperature probe or connectable temperature+humidity probe,
2x binary, counter, voltage and current input
Counted values
dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Memory
1 000 000 values (noncyclic record), 600 000 values (cyclic record)
Recording interval
1s to 24 hours
Display and alarm refresh
each 10 s
Recording mode
noncyclic – data logging stops after filling the memory
cyclic – after filling memory oldest data is overwritten by new
Available temperature units
degrees Celsius, degrees Fahrenheit
Communication interface
Ethernet, USB, RS232
Ethernet communication protocols
WWW, SNMPv1, XML, DATALINK
Alarm protocols
E-Mail
Power
AC adapter 230Vac/5Vdc (neccessary for Ethernet interface)
Battery life
3 – 4 weeks
Protection class
IP20
Dimensions (without connectors)
178 x 95 x 37mm, without attached cables
Weight (including batteries)
approx. 380 g
Warranty
3 years

  • code
    M1323
  • Reviews

    There are no reviews yet.

    Only logged in customers who have purchased this product may leave a review.