Multilogger – thermo hygro meter with 2 MiniDIN and 2 Thermocouple inputs

SKU: C95M1220 Categories: , , Tags: , , Brand:
BRAND
M1220

Universal datalogger with 4 inputs. Two inputs for temperature-humidity probes and two inputs for thermocouples K, J, S, B, T, N.

Multilogger can be installed permanently to Ethernet network or work as portable device.

Included is traceable calibration certificate in accordance with EN ISO/IEC17025.

The Multilogger is a battery operated datalogger capable of measuring and recording sensor and electrical values through its Inputs.
Data logger can be connected to Ethernet. Then measured values can be viewed on a web browser, as well as stored and downloaded for later analysis.

 

Technical data

Measured value


Temperature

Relative humidity

Counted values (dew point…)

TEMPERATURE SENSOR – external probe Pt1000

Measuring range
-200 to +600 °C
Accuracy
±0.2 °C (-200 to +100 °C);
±0.2 % from reading from +100 to +600 °C
Resolution
0.1 °C

TEMPERATURE+HUMIDITY SENSOR – external probe T+RH DIGI

Measuring range
according to the connected probe Digi/M
Accuracy
according to the connected probe Digi/M
Resolution
0.1 °C

DEW POINT

Measuring range
according to the connected probe Digi/M
Accuracy
according to the connected probe Digi/M
Resolution
0.1 °C

TEMPERATURE SENSOR – thermocouple K, N

Measuring range
-200 to +1300 °C
Accuracy
± 0,3 % from the measured value + 1,5 °C
Resolution
0.1 °C

TEMPERATURE SENSOR – thermocouple J

Measuring range
-200 to +750 °C
Accuracy
± 0,3 % from the measured value + 1,5 °C
Resolution
0.1 °C

TEMPERATURE SENSOR – thermocouple S

Measuring range
0 to +1700 °C
Accuracy
± 0,3 % from the measured value + 1,5 °C
Resolution
0.1 °C

TEMPERATURE SENSOR – thermocouple B

Measuring range
100 to +1800 °C
Accuracy
± 0,3 % from the measured value + 1 °C in the range +300 to +1800 °C
Resolution
0.1 °C

TEMPERATURE SENSOR – thermocouple T

Measuring range
-200 to +400 °C
Accuracy
± 0,3 % from the measured value + 1,5 °C
Resolution
0.1 °C
DC VOLTAGE
Measuring range
-60 to +140 mV; -18 to +18 mV
Accuracy
±100 µV; ±20 µV
Resolution
0,1 µV

GENERAL TECHNICAL DATA

Operating temperature
-10 to +60 °C
Channels
4x external temperature probe
Counted values
dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Memory
1 000 000 values (noncyclic record), 600 000 values (cyclic record)
Recording interval
1s to 24 hours
Display and alarm refresh
each 10 s
Recording mode
noncyclic – data logging stops after filling the memory
cyclic – after filling memory oldest data is overwritten by new
Available temperature units
degrees Celsius, degrees Fahrenheit
Communication interface
Ethernet, USB, RS232
Ethernet communication protocols
WWW, SNMPv1, XML, DATALINK
Alarm protocols
E-Mail
Power
AC adapter 230Vac/5Vdc (neccessary for Ethernet interface)
Battery life
3 – 4 weeks
Protection class
IP20
Dimensions (without connectors)
178 x 95 x 37mm, without attached cables
Weight (including batteries)
approx. 380 g
Warranty
3 years

  • code
    M1220
  • Reviews

    There are no reviews yet.